Customize Consent Preferences

We use cookies to help you navigate efficiently and perform certain functions. You will find detailed information about all cookies under each consent category below.

The cookies that are categorized as "Necessary" are stored on your browser as they are essential for enabling the basic functionalities of the site. ... 

Always Active

Necessary cookies are required to enable the basic features of this site, such as providing secure log-in or adjusting your consent preferences. These cookies do not store any personally identifiable data.

No cookies to display.

Functional cookies help perform certain functionalities like sharing the content of the website on social media platforms, collecting feedback, and other third-party features.

No cookies to display.

Analytical cookies are used to understand how visitors interact with the website. These cookies help provide information on metrics such as the number of visitors, bounce rate, traffic source, etc.

No cookies to display.

Performance cookies are used to understand and analyze the key performance indexes of the website which helps in delivering a better user experience for the visitors.

No cookies to display.

Advertisement cookies are used to provide visitors with customized advertisements based on the pages you visited previously and to analyze the effectiveness of the ad campaigns.

No cookies to display.

Duct Flange 4-20mA Dew Point humidity transmitter 20cm probe

Item Number:  HG803-3W4P-10

*The HG803 Air Duct Series Transmitter is equipped with a sensitive waterproof and anti fine dust high-temperature probe, with a temperature range of -40~80 ℃.
*Adopting standard industrial interface RS485 digital signal out put or 4-20 mA analog output, with display function, it can be connected to on-site digital display meters, PLCs, frequency converters, industrial control hosts an d other equipment.

 Modbus RS485 + 4-20mA Both Output

Product Highlights:

▪Integrated design, simple and elegant
▪Industrial grade ESD safety protection and power supply anti reverse connection design
▪Using waterproof, dustproof, and high-temperature resistant probes
▪Sensitive waterproof and anti fine dust high-temperature probe
▪Standard RS485 Modbus RTU communication protocol
▪High precision, high stability, LP66 waterproof and Dustproof level
▪Multiple probe materials available

SPECIFICATION

* Temperature Range: -40~80°℃(Plastic conjoined probe)
* Temperature Accuracy:土0.3°℃(25°C) ,(Higher precision is customizable)
* Humidity Range: 0~100% RH
* Humidity Accuracy:土3%RH(60%RH,25°C)(Higher precision is customizable)Long-term Temperature Stability: ≤0.1°C/y
* Long-term Humidity Stability: ≤1%RH/y
* Signal Output : ①The RS485 signal output or @4-20 mA temperature and humidity two analog capacity + RS485 signal output
* Baud rate: 1200、2400、4800、9600、19200、115200can set, Default is 9,600 pbs
* Transmitter circuit operating temperature: -20°C~+60℃,0%RH~95%RH,Non-knot dew Supports both single-register and multiple-register reads
* Industrial grade ESD safety protection and power supply reverse protection design

* Sensitive waterproof and fine dust high temperature resistant probe

*Digital output can set the output of 0.1 or 0.01 resolution

*Analog output with a resolution of 15 bits

HG803 Data Sheet Details

Technical DataDescription
DC power supply10 ~ 30V DC (12V recommended)
Maximum power consumptionAbout 1.2W
Current outputAbout 1.2W
Voltage outputAbout 1.2W
Typical accuracy (Temperature)±0.3℃ (25℃) / ±0.54℉ (77℉)
Typical accuracy (Humidity)±3%RH (60%RH, 25℃)
Measuring mediumAir
Measuring range (Temperature)-40℃ ~ 80℃ / -40℉ ~ 176℉ (default)
Measuring range (Humidity)0%RH – 100%RH
Long-term stability (Temperature)≤ 0.1℃ /y
Long-term stability (Humidity)≤ 1%RH /y
Response time (Humidity)≤ 8s (1m/s Wind speed)
Response time (Temperature)≤ 15s (1m/s Wind speed)
Analog outputSignal Current 4 – 20 mA
Digital outputRS485 Communication (Modbus) protocol
Data bit length8 bits
Parity check modeNone
Stop bit length1 bit
Default Modbus communication address1
Function code03
Baud rate1200, 2400, 4800, 9600, 19200, 115200 (optional, default 9600 bps)
Load capacity (Current output)≤ 600Ω
Load capacity (Voltage output)≤ 250Ω
Transmitter circuit operating temperature-20℃ ~ +60℃, 0%RH ~ 95%RH (non-condensation)

About Installation

HG803 Wire Connect Installation diagram

Wiring Standard of 4-20mA + RS485 Analog Output Type

Line ColorDefinitionSpecification
Power Cord  
RedPower supply positive10 ~ 30V DC
BlackPower supply negativeGND
Analog Output Lines  
BrownTemperature signal positive4-20mA output
BlueTemperature signal negativeConnects to Black (GND)
GreyHumidity signal positive4-20mA output
WhiteHumidity signal negativeConnects to Black (GND)
RS485 Wiring  
YellowRS485ATwisted pair cable
GreenRS485BTwisted pair cable

Additional Notes:

  • Load Capacity:
    • Current output: ≤ 600Ω
    • Voltage output (not applicable in this table): ≤ 250Ω
  • Applicable Models: HG803-1W4P, HG803-2W4P, HG803-3W4P, HG803-4W4P, HG803-5C4P, HG803-6C4P, HG803-7W4P, HG803-8W4P
 

Important:

  • Refer to the device manual for specific connection instructions and any additional requirements.
  • Ensure proper termination resistors are used on the RS485 bus as per the device manual and relevant communication standards.
  • Do not attempt electrical wiring without proper knowledge and qualifications.
  •  

APPLICATION

These transmitters can be used for indoor air temperature and humidity monitoring in various industrial plant, clean room, lab, machine room, commercial buildings, airport, and schools. 

The modbus interface is documented and integrator friendly. They also have transducer outputs for connecting as analog inputs to all popular control systems. 

Various models are available:

Duct Mount,

Wall Mount,Room Mount and 2 meter cable.BACnet MS/TP and 

Modbus RTU protocols over RS485,with an ethernet port.

Download Product Manual / Certification / Software

2. Check All Other Download Files

Blogs

Related Products

This is a gallery to showcase images from your recent social posts

Contact Us

en_USEnglish